Description
- Fast Response PRT, 3.2 mm (0.13 in), lead wire (cable) 6 ft
- 5624 Platinum Resistance Thermometer with Temperature Range of 0 °C to 1000 °C
- Temperature range of 0°C to 1000°C
- Accuracy (includes short-term stability and calibration uncertainty) of ±0.05°C to 962°C
- Long-term drift of 0.01°C at 0°C after 100 hours at 1000°C
- Designed by Fluke Calibration’s primary standards design team
Specifications
Range 0 °C to 1000 °C Calibration Uncertainty
± 0.004 °C at 0 °C
± 0.020 °C at 962 °CLong-term Drift (Rtpw) (k=2)
<0.01 °C at 0 °C/100 hours at 1000 °C
<0.06 °C at 0 °C/1000 hours at 1000 °CCalibrated Accuracy† (k=2) ± 0.006 °C at 0 °C
± 0.031 °C at 660 °C
± 0.041 °C at 961 °CShort-term Stability ± 0.005 °C Immersion <153 mm (6 in) at 700 °C Rtpw 10 ohm (± 0.1 ohm) Hysteresis <0.005 °C from 0 °C to 1000 °C Thermocycling <0.01 °C, 10 cycles from 0 to 1000 °C Current 1 mA Size 6.35 mm (0.25 in) O.D. Length 508 mm (20 in) Sheath Material Alumina Calibration Included 1913-6 fixed-point calibration